142 Chapter 8
P/N 130859-US, Manual Revision: AA Haemonetics
®
Cell Saver
®
Elite
®
+ User Manual
Volume By
Cycle Tab
The Volume By Cycle tab displays procedure statistics by cycle, including cycle
start time, duration, processed volume, wash volume, reinfusion volume,
concentrate volume, and for Sequestration PPP and PRP volume.
Figure 70, Example of the Volume By Cycle tab
If concentrate is used during the procedure, the volume of RBCs removed from
the RBC bag during the concentrate phase will be shown next to the reinfusion
volume for that cycle, as shown in Figure 70. The total reinfusion volume for
the procedure is equal to the sum of the reinfusion volume of each individual
cycle, minus the sum of the Conc volume for each cycle.