Troubleshooting 167
Haemonetics
®
Cell Saver
®
Elite
®
+ User Manual P/N 130859-US, Manual Revision: AA
104 Power Supply Failure
Explanation:
An issue has been detected with the power supply.
Corrective Action:
Restart the device.
If the problem persists, service is required.
105 Reservoir Sensor Failure
Explanation:
Auto-Fill is disabled. An issue has been detected with the reservoir sensor. Processing can
continue, but the device cannot automatically enter the Fill phase.
Corrective Action:
Touch Fill to start Fill when there is adequate fluid in the reservoir.
Touch OK to close this message.
Service is required to repair the reservoir sensor.
106 Waste Bag Sensor Failure
Explanation:
An issue has been detected with the waste bag sensor.
Corrective Action:
Restart the device.
If the problem persists, service is required.
107 Sensor Input Failure
Explanation:
An issue has been detected in one of the internal device components.
Corrective Action:
Restart the device.
If the problem persists, service is required.
Table 22, Event Messages