108 Air Detected in Yellow Line
Explanation:
The air detector sensed air in the yellow line. The saline bag may be empty.
Corrective Action:
1. Replace the saline bag if it is empty.
2. Ensure the yellow line clamp is open.
3. Check the tubing for kinks, occlusions, and proper placement in the air detector and pump.
4. Touch Continue to continue.
110 Air Detected During Conc
Explanation:
The air detector sensed air in the blue line during the Concentrate phase. The RBC bag may be
empty.
Corrective Action:
1. Ensure the blue line clamp is open.
2. Check the tubing for kinks, occlusions, and proper placement in the air detector and pump.
3. Wait for additional fluid in the reservoir.
OR
Tou ch Wash to wash a partial bowl.
111 Bowl is Ready to Wash
Explanation:
The optics have detected red cells. The bowl should be full and is ready to transition to the Wash
phase.
Corrective Action:
1. Touch Wash to enter the Wash phase OR
2. Touch to continue filling the bowl OR
3. Touch Return to return cells to the reservoir.
Table 22, Event Messages
ID # Event Message Text