IEC/EN 60601-1-2 Standard Requirements 221
Haemonetics
®
Cell Saver
®
Elite
®
+ User Manual P/N 130859-US, Manual Revision: AA
Table 27, Test specifications for enclosure port immunity to RF wireless communications
equipment
The system is intended for use in the electromagnetic environment specified below. The customer or
user of the system should ensure that it is used in such an environment.
Test
Frequency
(MHz)
Band
a)
(MHz)
Service
a)
Modulation
b)
Maximum
Power
(W)
Distance
(m)
Immunity
Test Level
(V/m)
385 380-390 TETRA 400 Pulse
modulation
b)
18 Hz
1.8 0.3 27
450 430-470 GMRS 460,
FRS 460
FM
c)
+5 kHz deviation
1 kHz sine
20.328
710 704-787 LTE Band 13,
17
Pulse
modulation
217 Hz
0.2 0.3 9
745
780
810 800-960 GSM 800/900,
TETRA 800,
iDEN 820,
CDMA 850,
LTE Band 5
Pulse
modulation
b)
18 Hz
20.328
870
930
1720 1700-1990 GSM 1800;
CDMA 1900;
GSM 1900;
DECT;
LTE Band 1, 3,
4, 25; UMTS
Pulse
modulation
b)
217 Hz
20.328
1845
1970
2450 2400-2570 Bluetooth;
WLAN,
802.11 b.g.n,
RFID 2450,
LTE Band 7
Pulse
modulation
b)
217 Hz
20.328
a) For some services, only the uplink frequencies are included.
b) The carrier shall be modulated using a 50% duty cycle square wave signal.
c) As an alternative to FM modulation, 50% pulse modulation at 18 Hz may be used, because while it does not represent actual modulation, it would be worst case.