Operation via fieldbus
100
a4643-1.0 HBM: public ClipX
Error 1-wire commu-
nication
12 The 1-wire TEDS cannot be read. Check the wiring. If
possible, check whether the TEDS module can be
read on another device, or is defective.
Error ClipX bus 13 The ClipX bus is not working correctly. Check the wir-
ing of the bus system.
Error external RAM 14 Error in the RAM of the ClipX (not in the RAM of the
CPU). Make a note of the error type and contact Tech-
nical support at HBM.
Error sensor exci-
tation
15 The excitation voltage for the sensor has been short-
circuited. Check the wiring of the sensor.
Fieldbus I/O 16 Cyclic communication is taking place on the fieldbus
(only BM40IE and BM40PB).
Fieldbus controller
defective
17 Internal error in the fieldbus controller (only on BM40IE
and BM40PB). Make a note of the error type and con-
tact
Technical support at HBM.
Error factory calibra-
tion
18 There is an error in the calibration of the ClipX. Make a
note of the error type and contact Technical support at
HBM.
Reserved 19
…
28
Not in use, and reserved for future upgrades.
Reading TEDS 29 The TEDS module is being read and the device is
being configured according to the specified settings.
Error external RAM 30 The data in the TEDS module either contain errors or
cannot be set.
Heartbeat 31
The bit is switched on and off rhythmically at 0.5Hz,
indicating that the ClipX is working.
Status/error Bit Explanation for bit set