7.1 General Specification
______________________________________________________________________________________________
4. Test signal levels
Open circuit voltage (V)
mode and
Constant voltage (CV)
mode
Range 10 mV to 5 V, 100 mA max. (1.000 MHz or less)
10 mV to 1 V, 20 mA max. (1.001 MHz or more)
Resolution 1 mV steps
Accuracy
10% 10 mV (1.000 MHz or less)
20% 10 mV (1.001 MHz or more)
Constant current (CC)
mode
Range 10 μA to 100 mA, 5 V max. (1.000 MHz or less)
10 μA to 20 mA, 1 V max. (1.001 MHz or more)
Resolution 10 μAsteps
Accuracy
10% 10 μA (1.000 MHz or less)
20% 10 μA (1.001 MHz or more)
Monitor function
Voltage monitor (Vmoni)
Range 0.000 V to 5.000 V
Accuracy
10% 10 mV (1.000 MHz or less)
20% 10 mV (1.001 MHz or more)
Current monitor (Imoni)
Range 0.000 mA to 100.0 mA
Accuracy
10% 10 μA (1.000 MHz or less)
20% 10 μA (1.001 MHz or more)
Limit function
Current limit (I-LIM)
(V, CV mode)
Range 0.01 mA to 99.99 mA
Accuracy
10% 10 μA (1.000 MHz or less)
20% 10 μA (1.001 MHz or more)
Voltage limit (V-LIM)
(CC mode)
Range 0.010 V to 5.000 V
Accuracy 10% 10 mV (1.000 MHz or less)
20% 10 mV (1.001 MHz or more)
5. Residual charge
protection
400 V max (for reference only)
Where:
{C: capacitance of the sample under test in farads}
6. Test ranges (RANGE)
The test range is determined according to impedance and phase angle. The
values of the other test parameters are calculated from |Z|.
Ranges
100 mΩ,1Ω,10Ω,100Ω,1kΩ,10kΩ, 100 kΩ,1MΩ,10MΩ,and
100 MΩ ( 10 ranges, can be set either automatically or manually)
|Z| display range
10.00 mΩ to 200.00 MΩ (5 digits)
θ
display range
+180.00 to -180.00 (5 digits)