161
_____________________________________________________________________________________________
7.1 General Specification
______________________________________________________________________________________________
1
2
3
4
5
6
7
8
9
10
11
12
13
14
A
7. Display ranges
The measurement range varies according to the test frequency.
( ): impedance
Y, G, B
99.999 S (100 mΩ) to 5.0000 nS (200 MΩ) 5 digits
R, X
10.00 mΩ to 200.00 MΩ 5digits
L
38.000 μH(100mΩ) to 750.00 kH(200MΩ) 5 digits
(at 42 Hz,
θ
=90 )
32.000 nH(10Ω) to 3.2000 mH(100kΩ) 5 digits
(at 5 MHz,
θ
=90 )
C
19.000 pF(200MΩ) to 370.00 mF(100mΩ) 5 digits
(at 42 Hz,
θ
=90 )
0.3200 pF(100kΩ) to 32.000 nF(10Ω) 5 digits
(at 5 MHz,
θ
=90 )
D
0.00001 to 9.99999 6 digits
Q
0.01 to 999.99 5 digits
NOTE: The test range of L and C varies depending on the test frequency
and value of θ.
8. Display
78.4 mm x 107.2 mm LCD display with backlight
9. Testing speed
Processing method: taken as the average of the waveform detected by the
measurement circuit.
When measuring open-
circuit terminal voltage
(V)
(minimum times with
test frequency 1 kHz,
Z display, initial screen)
The values depend on the measurement frequency, display parameter types,
open-circuit and short-circuit compensation, and whether or not the
comparator is being used.
FAST 5 ms
2ms
NORMAL 21 ms
2ms
SLOW 72 ms 2ms
SLOW2 140 ms
2ms
Constant voltage (CV)
and constant current
(CC) settings
Constant voltage/constant current measurement after connecting a
measurement sample
Maximum of 5 times measurement speed setting
10. Averaging function
Setting of the number
of averaging
OFF, 2, 4, 8, 16, 32, or 64
Processing
Internal trigger moving average of setting times
External trigger moving average of setting times
11. Trigger function
Manual setting of the Internal trigger (INT) or external trigger (EXT)
Trigger delay function (DELAY): 0.01 s to 9.99 s, setting resolution 0.01 s
12. Zero compensation
Open circuit
compensation (OPEN)
Correction of residual admittance between the measurement terminals of the
fixture
Operation is possible when, with the terminals open-circuit, the impedance
is at least 1 kΩ.
Short circuit
compensation (SHORT)
Correction of residual impedance between the measurement terminals of the
fixture
Operation is possible when, with the terminals short-circuit, the impedance
is less than 1 kΩ.
Spot compensation
(SPOT)/Continuous
correction (ALL)
For both open-circuit and short-circuit correction, a selection can be made
for correction either at the set frequency only, or correction to be made over
the entire frequency range.
13. Cable length setting
function
Canbesetto0mor1m