3.1 Pre-Operation Inspection
Touch the test lead to the zero-adjust board. If the
resistance display still shows a value of ‘- - . - -‘ the fuse
might be worn out or the test lead disconnected.
Is the test lead disconnected?
Is sufficient power remaining in the
battery?
The icon in the upper right hand area of the screen
indicates the current battery status. If the following
icon is displayed, the batteries are almost empty: Replace
the batteries.
Measurement may be impossible for insulated terminals,
due to an insufficient flow of current for measurement. In
such a case, clean the terminal (removing the insulation)
before measurement.
3.2 Setting the Measurement Range
Set resistance and voltage measurement ranges as described below.
(Single range) Since temperature measurement uses a signal range, range
setting is unnecessary.
Resistance Measurement Range
Press key to display the current settings. Press the key repeatedly to select the desired range.