3.3.1 Shorting Methods for Various Test Leads
Model 9465-10 Pin Type Lead
Short the test leads using the four terminal AC method, with the included zero-adjust board. As
shown in the illustration below, select a hole suited to the distance between terminals, on the
battery subject to measurement. Press it in a way symmetrical to the central screw on the zero-
adjust board. Keep the zero-adjust board at least 10cm away from the instrument.
Be sure to use the included zero-adjust board for the zero-adjust feature. Also, be
sure to connect each of the SOURCE and SENSE terminals by inserting the tip of
the pin into the holes on the zero-adjust board as shown above.
Do not place the zero-adjust board on top of the battery or any pieces of metal.
Electromagnetic induction effects could result in unstable measurement values. In
such a case, separate the zero-adjust board from any metal components.
Conducting zero adjustment by connecting the tips of the pin-type leads or using
a metal sheet other than the included zero-adjust board will result in inaccurate
adjustment.
When the distance between the terminals on the battery subject to measurement
is greater than the distance between the holes on the zero-adjust board, use the
holes in the corners for the zero-adjust feature.