4.3 Setting Comparator Permissible Values
Set the comparator permissible values (resistance upper limit no.1, resistance upper limit no.2 and
the voltage lower limit).
Example
Permissible values for a battery with initial values (i.e. resistance and voltage values when new or
in good condition) of 0.4Ω and 2V:
Resistance upper limit no.1: 0.6 Ω (1.5x initial value)
Resistance upper limit no.2: 0.8 Ω (2x initial value)
Voltage lower limit: 1.8V
Select the Comparator No.
1. Press key for two seconds or longer.
The comparator no. will flash. Press key again to
return to ordinary settings.
2. Press keys to select the comparator no. Press
keys to select digits.
3. Press this key to complete the setting of comparator nos.
and return to the range-setting screen.