(4) Enter resistance upper limit no.1 (R-Lim 1), resistance upper limit no.2 (R-Lim 2 ), and
the voltage lower limit (V-Lim).
Enter resistance input values in units of mΩ.
(5) Finalise the settings.
Configure all tables by repeating steps (1) – (5) above.
5. Save these configurations to a file by clicking [Save].
7.4.4 Editing Files of Permissible Values
1. Click [SET COMP/Clock] on the initialisation screen.
2. Click [COMP Table].
3. Click [Edit File].
4. Click comparator table file and then click [Open].
5. Editing configurations for each table.
Refer to ‘Section 7.4.3 Preparing a New Table of Permissible Values for how to set.