Industrial Batteries & Energy Storage
sales@regalelectro.com.au
A measurement value display of ‘OF’ indicates that the resistance or voltage
values displayed exceed the relevant measurement range.
If the voltage input if ‘OF’, a busser will sound.
A resistance value display of ‘- - - -‘ indicates that the test lead is open or
measurement cannot be conducted due to an electrical current irregularity
caused by a disconnected test lead or other reasons. ‘- - - -‘ will also be displayed
when the test lead is not correctly connected to the subject of measurement or
the subject’s resistance is substantially greater than the measurement range.
When measuring resistance of a relay or a connector, note the maximum open
circuit voltage of this instrument (approx. 5V). There is a possibility that such
measurement could damage the connectors of oxidised coating of the subject of
measurement, leading to incorrect measurement.
Ending the Measurement
Remove the test lead from the battery and turn off the power to the instrument by pressing
button.
3.7 Temperature Measurement
Use the optional 9460 Clip Type Lead with Temperature Sensor with temperature sensor to measure
battery temperature.
Connect the red connector of the clip-type lead with temperature sensor to the to the SENSE
terminal, the black connector to the SOURCE terminal, and the mini-plug to the TEMP.SENSOR
terminal. The instrument will detect the temperature sensor and display the measured temperature
automatically.
For safety reasons, when taking measurements, only use the optional 9460 Clip
Type Lead with Temperature Sensor with the instrument.
The sensor used in the temperature probe is a thin, precision platinum film. Be
aware that excessive voltage pulses or static discharges can destroy the film.