8.2 Accuracy
Accuracy Guarantee for
Temperature and
Humidity
23°C ± 5°C (73°F ± 9°F)
80%RH or less (non-condensating)
Guaranteed Accuracy
Conditions
Warming up: not required
Zero-Adjustment: Implemented
Temperature
Coefficient
Calculated using temperature coefficient
(T 23); T: Temperature used (°C)
18 to 28°C: no temperature coefficient
Resistance Measurement
Temperature Coefficient:
3mΩ range: (± 0.01 rdg. ± 0.8 dgt.)/ °C
Other range: (± 0.01 rdg. ± 0.5 dgt.)/ °C
Measurement current reliability: ± 10%
Measurement current frequency: 1kHz ± 30Hz
Range
Max. Displayed
Value
Resolution
Accuracy
Measured
Current
3mΩ
3.100mΩ
1µΩ
± 1.0% rdg. ± 8 dgt.
150mA
30mΩ
31.00mΩ
10µΩ
± 0.8% rdg. ± 6 dgt.
150mA
300mΩ
310.0mΩ
100µΩ
15mA
3Ω
3.100Ω
1mΩ
1.5mA
DC Voltage Measurement
Temperature Coefficient: (± 0.005 % rdg. ± 0.5 dgt.)/ °C
Range
Maximum
Displayed Value
Resolution
Accuracy
6V
±6.000V
1mV
± 0.08% rdg. ± 6 dgt.
60V
±60.00V
10mV