After approximately one second passes with no settings made, the settings displayed will be entered
and the instrument will return to the measurement screen.
Voltage Measurement Range
Press key to display the current settings. Press the key repeatedly to select the desired range.
After approximately one second passes with no settings made, the settings displayed will be entered
and the instrument will return to the measurement screen.
3.3 Zero-Adjustment
Using the zero-adjust feature makes more reliable measurements possible by adjusting the
resistance-range and voltage-range zero values of the instrument. Using the zero-adjust feature is
recommended prior to measurement. The zero-adjust feature sets the current measurement value
(adjusted value) at 0 and displays subsequent measurement results.
To prevent short circuits, do not place the zero-adjust board on top of the battery.
The zero-adjust feature takes approximately four seconds.
Using the zero-adjust feature adjusts the zero points of all ranges.
Even after turning off the power to the instrument, the adjust values will remain
and the zero-adjust feature will not be cancelled.
After replacing the test lead, be sure to use the zero-adjust feature prior to
measurement.
Be sure to use the included zero-adjust board for the zero-adjust feature.
Be sure to keep the test lead shorted during use of the zero-adjust feature. Keep
the tip of the test lead away from the metal components.