Measure Value: FAIL
Comparator Comparison Table
Judgement is conducted using the display and the buzzer, as shown in the following table:
Resistance Upper
Limit No.1
Resistance upper Limit
No.2
Boundary conditions are shown below:
Resistance
upper limit
no.1
Resistance
upper limit
no.2
Examples of How to Read the Comparator Output Table
When the measured resistance value does not exceed resistance upper limit no.1
and the measured voltage value is greater than the voltage lower limit, ‘Pass’ will be
displayed.
When the measured resistance value exceeds the resistance upper limit no.1 and
does not exceed the resistance upper limit no.2, and the measured voltage value is
greater than the voltage lower limit, ‘Warning’ will be displayed and a buzzer
sounded.