Industrial Batteries & Energy Storage
1300 734 253
sales@regalelectro.com.au
www.regalelectro.com.au
Operated Using
Formed rubber keys (18)
Display
LCD (monochrome, 159 segments)
Guaranteed Accuracy
Period
For one year
Operating
Temperature and
Humidity
0°C to 40°C (32°F to 104°F)
80%RH or less (non-condensating)
Storage Temperature
and Humidity
-10°C to 50°C (14°F to 122°F)
80%RH or less (non-condensating)
Operating Environment
Indoors, altitude up to 2000m (6562ft)
Power Supply
LR6 alkaline batteries x8
Rated Supply Voltage
DC1.5V x8
Maximum Rated Power
2VA
Continuous Operating
Time
Approx. 10 hours
When using alkaline batteries; may vary depending on conditions of use
Dimensions
Approx. 192 (W) x 121 (H) x 55mm (D)
Mass
Approx. 790g
Dielectric Strength
Between all measurement terminals and USB terminal:
AC 1.5kV; 15 seconds; cut-off current: 5mA
Maximum Input
Voltage
Between positive and negative measurement terminals:
DC 60V
Maximum Rated
Voltage to Earth
Between all measurement terminals and grounds:
DC 70V
Applicable Standards
EMC
EN61326:1997+A1:1998+A2:2001+A3:2003 Safety
EN61010-1:2001 Pollution degree 2 Measurement category 1
(anticipated transient overvoltage 500V)
Effect of Radiated
Radio-Frequency
Electromagnetic Field
± 3.0%f.s. when measuring both resistance and voltage (at 3V/m)