EasyManua.ls Logo

Hioki DT4261 - AC+DC Current Specs; Troubleshooting

Hioki DT4261
154 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
125
Input and Measurement Specications
16. AC + DC current
RMS value (Measured value/MAX/MIN/AVG)
Range
Display range
(Accuracy
guarantee range)
Accuracy*
1
Input
impedance
DC, 40 Hz ≤ f
≤ 500 Hz
500 Hz < f ≤
1 kHz
600.0 mA
0.0 mA to 600.0 mA
(6.0 mA to 600.0 mA)
±1.4% rdg
±3 dgt
±1.8% rdg
±3 dgt
35 m
Ω
±30%6.000 A
0.000 A to 6.000 A
(0.060 A to 6.000 A)
±1.4% rdg
±3 dgt
±1.8% rdg
±3 dgt
10.00 A
0.00 A to 10.00 A
(0.10 A to 10.00 A)
±1.4% rdg
±3 dgt
±1.8% rdg
±3 dgt
Crest factor: 3 up to 4000 counts
Reduces linearly to 2 at 6000 counts (other than 10.00 A range).
1.5 or less (10.00 A range)
Coupling type: DC coupling
Auto range movement threshold: More than 6000 counts for upper range
Less than 540 counts for lower range
*1: ±5 dgt should be added to 5% or less of the range.
When the lter is ON
100 Hz: ±1.5% rdg added in the range of 40 Hz to 100 Hz,
no specied accuracy over 100 Hz
500 Hz: ±0.5% rdg added in the range of 40 Hz to 500 Hz,
no specied accuracy over 500 Hz
Peak value (PEAK MAX/PEAK MIN)
Range
Display range
(Accuracy
guarantee range)
Accuracy
DC, 40 Hz ≤ f ≤ 500 Hz 500 Hz < f < 1 kHz
600.0 mA
−1200 mA to 1200 mA
(±30 mA to ±1200 mA)
±1.5% rdg ±7 dgt ±2.0% rdg ±7 dgt
6.000 A
−12.00 A to 12.00 A
(±0.30 A to ±10.00 A)
±1.5% rdg ±7 dgt ±2.0% rdg ±7 dgt
10.00 A
−15.00 A to 15.00 A
(±0.50 A to ±10.00 A)
±1.5% rdg ±7 dgt ±2.0% rdg ±7 dgt
Input impedance, coupling type:
The same as the AC + DC current RMS value
Range movement:
Based on the range movement of the AC + DC current RMS
value
HIOKI DT4261A961-00

Table of Contents

Other manuals for Hioki DT4261

Related product manuals