124
Input and Measurement Specications
14. AUTO A (AC current/DC current automatic judgment)
AC judgment: The same as the accuracy specications in “16. AC + DC
current” (p. 125)
DC judgment: The same as the accuracy specications in “15. DC current”
(p. 124)
15. DC current
Measured value (Measured value/MAX/MIN/AVG)
Range
Display range
(Accuracy guarantee
range)
Accuracy*
1
Input
impedance
600.0 mA −600.0 mA to 600.0 mA ±0.5% rdg ±3 dgt
35 m
Ω
±30%6.000 A −6.000 A to 6.000 A ±0.5% rdg ±3 dgt
10.00 A −10.00 A to 10.00 A ±0.5% rdg ±3 dgt
Coupling type: DC coupling
Auto range movement threshold: More than 6000 counts for upper range
Less than 540 counts for lower range
*1: ±2 dgt should be added to 5% or less of the range.
Peak value (PEAK MAX/PEAK MIN)
Range
Display range
(Accuracy guarantee
range)
Accuracy
Input
impedance
600.0 mA −1200 mA to 1200 mA ±1.5% rdg ±7 dgt
35 m
Ω
±30%6.000 A −10.00 A to 10.00 A ±1.5% rdg ±7 dgt
10.00 A −10.00 A to 10.00 A ±1.5% rdg ±7 dgt
Coupling type: The same as the DC current measured value
Range movement: Based on the range movement of the DC current
measured value
HIOKI DT4261A961-00