78
Hold Function (HOLD)
Requirements for auto holding
When the following 2 requirements are met, display update is
stopped.
• When the uctuation width of the measured value stabilizes within
the range shown in the following table
• When the measured value exceeds the threshold value shown in
the following table (voltage, current) or the measured value falls
below the threshold value in the following table (continuity check,
resistance, diode test)
Measurement
item*
1
Range Fluctuation range
Threshold
value
AUTO V*
2
DC voltage*
2
AC + DC voltage
AC voltage
LoZ V
Other than
1000 V
Within 120 counts 120 counts
1000 V Within 20 counts 20 counts
Continuity check
Resistance
All ranges Within 100 counts 4900 counts
Diode test 1.800 V Within 40 counts 1460 counts
AC current
(Clamp sensor)
10.00 A Within 50 counts 50 counts
20.00 A Within 100 counts 100 counts
50.0 A Within 25 counts 25 counts
100.0 A Within 50 counts 50 counts
200.0 A Within 100 counts 100 counts
500 A Within 25 counts 25 counts
1000 A Within 50 counts 50 counts
AC current
AUTO A
DC current
AC + DC current
Other than 10 A Within 120 counts 120 counts
10 A Within 20 counts 20 counts
*1: Auto holding is not available for measurement items not shown.
*2: Auto holding is not available for the 600 mV range.