Chapter 1. Overview
1-3
Chapter 1.
(3) When using the cartridge bottle makeup as a cleaning solution
(a) Open the makeup cartridge bottle by turning its cap.
At this time, do not remove the over cap.
(b) Use the makeup by transferring it to the cleaning bottle.
[Note]
If no cleaning solution is available, transfer the makeup in the makeup cartridge bottle to the
cleaning bottle before setting the makeup cartridge bottle.
Tighten the cap of the makeup cartridge bottle after use. Turn it upside down to check if there is
no leakage of makeup.
If the makeup in the makeup cartridge bottle is used as a cleaning solution, the volume of
makeup in the makeup cartridge bottle decreases. When it is set in the unit, please use a makeup
cartridge bottle that contains makeup more than half of the bottle.
If makeup is accidentally spilled, promptly wipe it up with wiping paper.
Cap
Over cap
Makeup
cartridge bottle
Cleaning
bottle