Page 166 ST 700 SmartLine Pressure Transmitters User’s Manual Revision 13
Barometric
Correct
[27BARCOR]
One point field offset
Calibration Method.
Available for AP models only.
Executing this selection
corrects input pressure to the
Barometric Standard value
when the pressure input is
open to the atmosphere.
The live value of the pressure
input is shown on this display
so the user can easily see the
result of the correction.
Press to
enter menu
selection
Press to
initiate action
Calibrate High
Point
[28CAL HI]
The upper pressure value for a
two-point field calibration within
the LRV to URV span.
Cal high point value in inH2O
[4°C]
Range: LRV to URV
Default: URV
Press to
enter menu
selection
Menu to
select
number.
to enter
and shift to
the next digit
to the right
Calibrate Low
Point
[29CAL LO]
The lower pressure value for a
two-point field calibration within
the LRV to URV span. Cal low
point value in inH2O [4°C]
Range: LRV to URV
Default: LRV
Press to
enter menu
selection
Menu to
select
number.
to enter
and shift to
the next digit
to the right
Calibrate Adjust
Ratio
[30CALRTO]
A ratio value that adjusts the
calibrated curve between the
Cal Low and High Points.
Error Adjustment Ratio
Range: 0.5 to 1.5
Default: 1.0
Press to
enter menu
selection
Menu to
select
number.
to enter
and shift to
the next digit
to the right