Use the HP 54713B plug-in when:
The signal happens once or infrequently, and you are using the real-time
sampling mode.
Your application requires only 500-MHz bandwidth, (rise time is slower
than 2.1 ns) and you need more than 2 channels.
You application requires a 1 M
input because you are using a
passive-compensated divider probe.
Use the HP 54714A plug-in when:
Your application requires up to 8 channels.
Use the HP 54721A plug-in when:
The signal happens once or infrequently, and you are using the real-time
sampling mode.
You need the 1.1-GHz real-time bandwidth (rise time is slower than 1 ns).
You need the 4-GSa/s sample rate.
You need 32K memory depth.
Use the HP 54722A plug-in when:
The signal happens once or infrequently, and you are using the real-time
sampling mode.
You require the 8-GSa/s sample rate.
You need high, real-time bandwidth.
Your application requires high trigger bandwidth, and you can use external
triggering.