USER CAL
9
This softkey initiates a full synthesizer user calibration. The
calibration performed is instrument state dependent. For example, if
the synthesizer is in ramp sweep mode, a sweep span calibration and
an auto track is done. If the synthesizer has amplitude modulation
active on a CW signal, then RF peaking and an AM bandwidth
calibration is performed.
SCPI:
See the individual types of calibration.
Analyzer: NONE
AM BW Cal Always,
AM BW Cal
Once,
Auto Track,
Peak RF Always,
Peak RF Once, Swp Span Cal Always,
Swp Span Cal Once