l
DIAGnostics:TEST:LOG[:STATe]?
l
DIAGnostics:TEST:LOG[:STATe] ON|OFF|l|O
Selects and queries the raw data logging ON/OFF switch. Both
commands are executable in
selftest
mode.
After
*RST,
the setting is 0.
l
DIAGnostics:TEST:LOOP ON|OFF|l|O
l
DIAGnostics:TEST:LOOP?
Selects and queries the test looping ON/OFF switch. Both
commands are executable in
selftest
mode.
After “RST, the setting is 0.
l
DIAGnostics:TEST:NAME?
[<num>]
Queries the name of a
selftest
by number. If the number is not
specified then an array of all the
selftest
names is returned.
l
DIAGnostics:TEST:POINts?
Returns the number of points of
selftest
that is output using
DIAGnostics:TEST:NAME? or DIAGnostics:TEST:RESult?.
l
DIAGnostics:TEST:RESult?
[<num>]
Queries the result of a selftest, by number. The response is a string
containing either Passed, Failed, or
NotRun.
If <num> is missing,
an array of
selftest
results are returned.
l
DIAGnostics:TINT?
<num>
A test feature that returns the value passed to it. This is used to test
the HP-IB interface.
l
DISPlay[:STATel
ON|OFF|l|O
l
DISPlay[:STATe]?
Sets and queries the display ON/OFF switch.
After
*RST,
the value is 1.
l
FM:COUPling
ACiDC
l
FM:COUPling?
Sets and queries the FM input coupling mode.
The *RST value is AC.
l
FM[:DEViation]
<num>[freq
suffix>llMAXimumlMINimum
l
FM[:DEViation]?
[MAXimumlMINimum]
Sets and queries the peak FM deviation (in Hz) when the internal
FM generator is used. After
'RST,
the value is 1 MHz.
l
FM:FILTer:HPASs
<num>[freq
suffix'lMAXimumlMINimum
l
FM:FILTer:HPASs?
[MAXimumlMINimuml
Sets and queries the FM AC bandwidth. There are only two
S-34