2
This
softkey
lets you define the number of step points in a stepped
frequency sweep. The number of points in a stepped sweep can
range from
2 to 801. Step Size and Step Points are dependent
variables. If you know how many steps are desired in a given sweep,
use the
softkey
Step Points to set the desired value The step size
will be calculated automatically.
SCPI: SWEep[:FREQuency]:POINts
<num>]MAXimum]MINimum
Analyzer: NONE
Step Size, Step Swp Menu, Sweep Mode Step
“Using Step Sweep,” in Chapter 1.
Step Size
Function Group
FREQUENCY
Menu Map
2
Description
This
softkey
lets you specify the step size in a stepped frequency
sweep.
The range of increment size is dependent on frequency
span and the number of step points desired, as given by the
formula: STEP SIZE = SPAN
+
STEP POINTS. Step Size and
Step Points are dependent variables, as shown by the formula. If
a particular step size is desired, use the Step Size
softkey
to set
the desired increment. The number of step points is then calculated
automatically.
Programming Codes
SCPI: SWEep[:FREQ
uency]:STEP
<num>[freq suffix] or
MAXimumjMINimum
Analyzer: NONE
See Also
Step
Points, Step Swp Menu, Sweep
Mode Step
“Using Step Sweep,” in Chapter 1.