EasyManua.ls Logo

Inficon IC6

Inficon IC6
342 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5 - 18
IPN 074-505-P1E
IC6 Operating Manual
CAUTION
The Sensor must be positioned so the rate can be
monitored with the source shutter closed when going to
the non-deposit control state.
RAMP 1 RATE. . . . . . . . . . . . . . . . . 0.000 to 999.9 Å/s
Rate Ramp 1 is provided to effect an aggregate rate change while depositing a
Layer. The aggregate rate is ramped linearly from the original rate setpoint to
the New Rate 1 value over the time interval Ramp 1 Time. It is allowed to ramp
to a higher or a lower rate. The default value is 0.0.
START RAMP 1. . . . . . . . . . . . . . . . 0.000 to 999.9 kÅ
This value sets the thickness at which to start the Rate Ramp. This thickness
value must be less than the Start Ramp 2 thickness value if the value of Start
Ramp 2 is non-zero. A value of zero disables the feature. The default value is 0.
RAMP 1 TIME . . . . . . . . . . . . . . . . . 00:00 to 99:59 min:s
This value determines the time period over which to ramp the rate from the
original rate to New Rate 1. The default value is 00:00.
RAMP 2 RATE. . . . . . . . . . . . . . . . . 0.000 to 999.9 Å/s
Ramp 2 Rate is provided to effect an aggregate rate change while depositing a
Layer. The aggregate rate is ramped linearly from the original Ramp 1 Rate
setpoint to the New Rate 2 value over the time interval Ramp 2 Time. It is
allowed to ramp to a higher or a lower rate. The default value is 0.0.
START RAMP 2. . . . . . . . . . . . . . . . 0.000 to 999.9 kÅ
This value determines the thickness at which to begin Ramp 2 Rate. A value of
zero disables this feature. Start Ramp 2 thickness must be greater than Start
Ramp 1 thickness. Also, Ramp 2 Rate can not be started until the completion
of Ramp 1 Rate. If the thickness of Start Ramp 2 is exceeded while in Ramp 1
Rate, Ramp 2 Rate will begin immediately after Ramp 1 Rate The default value
is 0.
RAMP 2 TIME . . . . . . . . . . . . . . . . . 00:00 to 99:59 min:s
This value is the time interval over which the aggregate rate is ramped linearly
from New Rate 1 to New Rate 2. The default value is 00:00.

Table of Contents

Related product manuals