15 - 18
IPN 074-505-P1E
IC6 Operating Manual
6. Control voltage output
does not function properly
a. DAC board damaged from
applying voltage to the
control voltage output
a. Ensure cable connection
to the DAC board does not
have a potential across the
contacts, contact INFICON
service department
b. Reversed polarity of
control voltage relative to that
accepted by the source
power supply
b. Verify source output
polarity of the DAC and the
required input polarity of the
source power supply, refer to
the instruction manual to
reconfigure the IC6 if
necessary
c. Improper control cable
fabrication
c.Check for correct cable
wiring in the appropriate
section of the manual
7. LCD display dull or blank a. Brightness/contrast
adjustment required
a. Refer to manual for
location of adjustment
potentiometer, adjust as
desired
b. LCD or power supply
problem
b. Contact INFICON service
department
8. Poor rate control a. Control loop parameters
improperly selected
a. Refer to the instruction
manual section on tuning
control loop parameters
b. Electron beam sweep
frequency "beating" with the
IC6's measurement
frequency
b. Adjust the sweep
frequency so it is not a
multiple of the IC6's
measurement frequency
Table 15-7 Troubleshooting the IC6 (continued)
SYMPTOM CAUSE REMEDY