28 Installation
iina74e 02.fm technical handbook(1408)
2.5 Default parameters
The following parameters are set like shown when in the menu of the UL5000 under
Settings Parameters saveload, „load default values“ is chosen.
Auto-scaling: On
Scaling logarithmic
Display range: 4 decades
Time axis: 32 seconds
LCD invers: OFF
Background in stand by mode: OFF
Automatic calibration request: OFF
Mass: 4 (helium)
Recorder Output: leak rate
Volume: 2
Zero time: 5 seconds
Leak rate unit: mbar l/s
Mode: Vacuum with
HYDRO•S
Trigger level 1: 1×10
-9
mbar l/s
Trigger level 2: 1×10
-8
mbar l/s
Leak rate external test leak (Vacuum): 1×10
-7
mbar l/s
Leak rate external test leak (Sniffer): 1×10
-5
mbar l/s
Vent delay: 2 seconds
Automatic purge: OFF
Pressure unit: mbar
Minimum volume: 0
Beep: ON
Maximum evacuation time: 30 minutes
Audio Alarm Type: Trigger Alarm
Maximum inlet pressure when sniffing: 1 mbar
Minimum Inlet pressure when sniffing: 0,1 mbar
Number of decimal place at leak rate displayed: 1
Scroll display: On
Particle protection: Off
Direct access to calibration: On
Contamination protection: Off
Switch off limit for contamination protection: 1E-3 mbar l/s
Control location: Local
Alarm delay: 30 seconds
Leak rate filter: I•Cal
Zero: enable
HYDRO•S automatic enabled