58 Description of the Menu
iina74e 06.fm technical handbook(1408)
6.6.1 Vacuum settings
• Main Menu > Settings > Vacuum settings
This menu allows to observe and to change the settings belonging to the vacuum
system.
6.6.1.1 Automatic purge
• Main Menu > Settings > Vacuum settings > Automatic purge
Through this menu it is possible to program the automatic purge (
Please refer to
Chapter
5.3.1) for 20 seconds when switching from measuring to standby mode.
Softkey 2: Automatic purge (Purging)
Refer to Chapter
6.6.1.1
Softkey 3: Vent delay
Please refer to Chapter 6.6.1.2
Softkey 4: Booster TMP mode
Please refer to Chapter 6.6.1.7
Softkey 5: Vacuum ranges
See Chapter 6.6.1.3
Softkey 6: Leak rate internal test leak
Please refer to Chapter 6.6.1.5
Softkey 7: Machine factor
Please refer to Chapter 6.6.1.6
Softkey 8: Booster TMP mode
Please refer to Chapter 6.6.1.7
Softkey 3: OFF
Automatic purge is switched off at standby mode.
Softkey 7: ON
Automatic purge is activated. When switching from measurement to STAND-
BY the forepump is rinsed automatically for 20 seconds.