EasyManuals Logo

Inficon UL5000 User Manual

Inficon UL5000
108 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #71 background imageLoading...
Page #71 background image
Description of the Menu 71
iina74e 06.fm technical handbook(1408)
6.6.6.1 Save parameter set
Main Menu > Settings > Parameter save / load > Save
Save the current parameter settings.
6.6.6.2 Load parameter set
Main Menu > Settings > Parameter save / load > Save parameter set
The settings of the
selected saved parameter set will be displayed and can be
reloaded.
6.6.7 Monitoring functions
Main Menu > Settings > Monitoring functions
Calibration request
Main Menu > Settings > Monitoring functions > Calibration request
It can be selected whether the operator is reminded of the fact that a calibration may
has become necessary or not. The default value is off.
If the calibration request is switched on, a corresponding message will appear when
30 minutes have elapsed after power on or if the temperature of the UL5000 has
changed by more than 5 °C (9 °F) since the last calibration.
Particle protection
Main Menu > Settings > Monitoring functions > Particle protection
This mode can be switched on and off.
If switched on the UL5000 will not start pumping down before the inlet ha
s not
dropped below 1 mbar. I. e. it is assumed that the part under test is pumped by
another pump in parallel.
Purpose: When the leak detector does not pump at high pressures no gas stream
,
possibly carrying particles gets into the leak detector.
Softkey 4: Edit a file name
Rename the parameter set.
Softkey 6: Arrow up
Upward to the previous screen.
Softkey 7: Arrow down
Downward to the next screen.
Softkey 3: Off
The calibration request will be switched off.
Softkey 7: ON
The calibration request will be switched on.