Error And Warning Messages 87
iina74e 08.fm technical handbook(1408)
W83 All EEPROM parameter lost.
Please check your settings.
• EEPROM on back plane is empty and was initialized with default
valves. Enter all parameters again.
• The EEPROM might be faulty when warning comes up again after
power up.
W85 Lost EEPROM parameter!
Please check your settings!
• Writing access was interrupted. Please check all adjustments.
• An update of software was done. In this case the notice can be ignored.
• When warning comes up again after powering up the EEPROM might
be faulty.
W86 AC/DC factor too low Calibration conditions not maintained
Value of leak rate not entered correctly
Test leak faulty
W87 AC/DC factor too high Calibration conditions not maintained
Value of leak rate not entered correctly
Test leak faulty
No. Displayed Message Description and possible solutions