Seraâ„¢ - Instructions for Use - EN Page 78
Guidance and Manufacturer’s Declaration - Electromagnetic Immunity
The Sera
TM
is intended for use in the electromagnetic environment specif ied below. The customer or the
user of the Sera
TM
should ensure that it is used in such an environment.
Electromagnetic
Environment-Guidance
Discharge (ESD)
IEC 61000-4-2
+15 kV air
+15 kV air
concrete, or ceramic tile. If
floors are covered with
synthetic material, the relative
humidity should be greater
transient/burst
IEC61000-4-4
lines
+1 kV for input/output
lines
supply lines
+1 kV f or input/output
lines
Mains power quality should be
that of a typical commercial or
residential environment.
IEC 61000-4-5
+2 kV common mode
mode
+2 kV common mode
Mains power quality should be
that of a typical commercial or
residential environment.
interruptions, and
voltage variations on
power supply lines
IEC 61000-4-11
(>95% dip in UT) f or 0.5
cycle
40% UT
(60% dip in UT) f or 5
cycles
70% UT
(30% dip in UT) f or 25
cycles
<5% UT
(>95% dip in UT) f or 5
UT)
for 0.5 cycle
40% UT (60% dip in
UT) f o r
5 cycles
70% UT (30% dip in
UT) f o r
25 cycles
<5% UT
Mains power quality should be
that of a typical commercial or
residential environment. If the
user of the Sera
TM
requires
continued operation during
power mains interruptions, it is
recommended that the Sera
TM
be powered from an
uninterruptable power supply
or its battery.
(50/60 Hz)
IEC 61000-4-8
Power f requency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or residential
Note: UT is the A.C. mains voltage prior to application of the test level.