D-0112389-I – 2022/04
Sera™ - Instructions for Use - EN Page 79
Guidance and manufacturer’s declaration — electromagnetic immunity
The Sera
TM
is intended for use in the electromagnetic environment specif ied below. The customer or the
user of the Sera
TM
should assure that it is used in such an environment,
IEC / EN 60601
test level
Electromagnetic environment
– guidance
Conducted RF
IEC / EN 61000-4-6
Radiated RF
IEC / EN 61000-4-3
3 Vrms
150kHz to 80 MHz
3 V/m
80 MHz to 2,7 GHz
3 Vrms
3 V/m
communications equipment
should be used no closer to
any parts of the Sera
TM
,
including cables, than the
recommended separation
distance calculated from the
equation applicable to the
frequency of the transmitter.
Recommended separation
distance
80 MHz to 800
MHz
w
800 MHz to
2,5 GHz
Where P is the maximum
output power rating of the
transmitter in watts (W)
according to the transmitter
manufacturer and d is the
recommended separation
distance in meters (m).
Field strengths f rom fixed RF
transmitters, as determined by
an electromagnetic site
survey, (a) should be less than
the compliance level in each
frequency range (b)
Interf erence may occur in the
vicinity of equipment marked
with the following symbol: