EasyManuals Logo

ITech IT-M7700 Series Programming Guide

ITech IT-M7700 Series
126 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #37 background imageLoading...
Page #37 background image
IEEE-488 Common Commands
Copyright© Itech Electronic Co., Ltd. 25
Related Command
*ESE *ESR? *PSC *STB?
*STB?
Status byte query. The Status Byte is a read-only register and the bits are not
cleared when it is read.
Query Syntax
*STB?
Arguments
None
Returns
<NR1>
Related Command
*CLS *ESE *ESR
The following table describes the Status Byte Register register bit assignments.
Bit
7
6
5
4
3
2
1
0
Name
OPER
RQS
ESB
MAV
QUES
EAV
n.u
n.u
Decimal
Value
128
64
32
16
8
4
*PSC
This command controls whether a service request will generate when the
power supply is re-generalized.
1 OR ON: when the power supply is generated, the values in the Status
Byte Enable register, Operational Event Enable register, Query Event
Enable register and Standard Event Enable register are cleared.
0 OR OFF: the values in the Status Byte Enable register, Operational Event
Enable register, Query Event Enable register and Standard Event Enable
register are saved in NVM, which will be used at regenerating.
Syntax
*PSC <ON|1|OFF|0>

Table of Contents

Other manuals for ITech IT-M7700 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ITech IT-M7700 Series and is the answer not in the manual?

ITech IT-M7700 Series Specifications

General IconGeneral
BrandITech
ModelIT-M7700 Series
CategoryPortable Generator
LanguageEnglish

Related product manuals