LIST Subsystem(IT-M7700)
Copyright© Itech Electronic Co., Ltd. 61
Chapter 9 LIST Subsystem(IT-M7700)
LIST commands is a special command of
IT-M7721/IT-M7722/IT-M7723/IT-M7723E, but it’s not suitable for
IT-M7721L/IT-M7722L.
LIST:TRIGger:MODe
This command is used to set the trigger mode to run the List.
Syntax
LIST:TRIGger:MODe <OFF|ON|0|1>
Arguments
<OFF|ON|0|1>
Off: After the Trigger key is pressed, the instrument will automatically run
the real-time List file.
On: After the Trigger key is pressed, the instrument will run the realtime
List file in single step.
Query Syntax:
LIST:TRIGger:MODe?
Returns:
<OFF|ON>
LIST:STATe
This command is used to set the mode of recall list.
Syntax
LIST:STATe <OFF|1-5>
Arguments
<OFF | 1-5>
1 - 5: indicate that the List Mode function is enabled, and LIST1 to LIST5
file is called.
Off: indicating that the List Mode function is disabled.