Model 4200A-SCS Pulse Card (PGU and PMU) User's Manual Section 4: Pulse card concepts
4200A-PMU-900-01 Rev. A December 2020 4-9
“Pulsed sourcing” can involve outputting user-defined two-level pulses, outputting multi-level pulses
using the built-in Segment Arb
®
function, or outputting an arbitrarily defined waveform using the
arbitrary waveform generator in the KPulse software, which is included with the 4200A-SCS. The
Segment Arb feature allows you to create waveforms from segments defined with separate voltages
and time durations. In addition to its pulse generator capabilities, the 4225-PMU can measure ac or
dc voltage and current, thereby reducing the need for additional measurement hardware and more
complicated programming. The 4220-PGU is a two-channel pulse generator with output capabilities
identical to the capabilities of the 4225-PMU, but without its measurement functions.
Sample rate
For the PMU, the maximum measurement sampling rate for each A/D test is 200e6 (200 million)
samples per second. However, there is a limit to the number of samples (one million) that can be
acquired per A/D test. When a test is configured to exceed that limit, the sample rate is automatically
lowered when using ITMs so that fewer than one million samples are acquired.
Pulse I-V (spot mean)
The maximum number of samples per A/D per test is <1,000,000 (one million). If an ITM is configured
to yield more than one million samples, Clarius automatically lowers the sampling rate. For pulse I-V
(spot mean), typically the sample rate is reduced only if the measure window is wide (due to a wide
pulse width or period) or if the number of pulses is large.
The total number of samples for a test is calculated as follows:
Number of samples = Measure window x Sample rate x Number of pulses x Sweep points x Step
points
Example: Test that uses a single PMU to perform 50 20-step sweeps
Pulse width = ~7 µs
Measure window = 1 µs
Sample rate = 200e6 samples per second
Number of pulses = 50
Number of steps in sweep = 20
The number of samples acquired for the above example is calculated as follows:
Number of samples = 1 µs x 200e6 x 50 x 20 = 200,000
Because the number of samples is less than the one million sample limit, the sample rate of 200e6
samples per second is used for the above example.

Table of Contents

Other manuals for Keithley 4200A-SCS

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 4200A-SCS and is the answer not in the manual?

Keithley 4200A-SCS Specifications

General IconGeneral
BrandKeithley
Model4200A-SCS
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals