34410A/11A/L4411A User’s Guide 137
Specifications 5
Temperature Characteristics
Accuracy Specifications
[1]
[ 1 ] For total measurement accuracy, add temperature probe error
Examples (RTD probe, measurement within “Best Range”):
1.) Meter within T
CAL
± 5 °C: Error = 0.06
o
C + probe error
2.) Meter at T
CAL
+ 10 °C: Error = 0.06
o
C + (5 x 0.003
o
C) + probe error = 0.075
o
C + probe
error
Function Probe Type R
o
Best Range 24 Hour/
90 Day/1 Year
T
CAL
± 5 °C
Temperature
Coefficient
0 °C to (T
CAL
– 5 °C)
(T
CAL
+ 5 °C) to 55 °C
Temperature RTD
Thermistor
from 49 Ω to 2.1 kΩ
N/A
–200
o
C to 600
o
C
–80
o
C to 150
o
C
0.06
o
C
0.08
o
C
0.003
o
C
0.002
o
C

Table of Contents