86 34410A/11A/L4411A User’s Guide
2 Features and Functions
Refer to the Keysight 34410A/11A/L4411A Programmers Reference Help for
a complete description and syntax for this command.
NISPOM Memory Sanitization
You can sanitize all user-accessible instrument memory either from the front panel, or
from the remote interface. This complies with the requirements in chapter 8 of the
National Instrument Security Program Operating Manual (NISPOM).
Front Panel Operation: Press
UTILITY MENU > NISPOM
Select “YES” to sanitize all user-accessible memory.
Remote Interface Operation: The following command sanitizes all user-accessible
instrument memory.
SYSTem:SECurity:IMMediate
Refer to the Keysight 34410A/11A/L4411A Programmers Reference Help for
a complete description and syntax for this command.
Calibration
Only service–qualified personnel should calibrate the multimeter. Improper use of the
front–panel CALIBRATION procedure in the UTILITY menu could result in damage to
the multimeter. The multimeter is secured with a calibration security code at the factory.
Refer to the Keysight 34410A/11A/L4411A Service Guide for calibration procedures.