34410A/11A/L4411A User’s Guide 67
Features and Functions 2
The Limits annunciator indicates that the limits function is enabled. If the reading is
lower than the low limit, LO will flash. If the reading exceeds the high limit, HI will
flash. In either case, when a limit is crossed, the multimeter will also beep, if the
beeper is enabled.
Remote Interface Operation: The following command sequence selects the limit
test function, turns that function on, and sets the lower and upper limits. You can
assign a lower limit, an upper limit, or both.
CALCulate:FUNCtion LIMit
CALCulate:STATe ON
CALCulate:LIMit:LOWer <
value
>
CALCulate:LIMit:UPPer <
value
>
The following commands query the selected lower and upper limits.
CALCulate:LIMit:LOWer?
CALCulate:LIMit:UPPer?
Refer to the Keysight 34410A/11A/L4411A Programmers Reference Help for
a complete description and syntax for these commands.