34410A/11A/L4411A User’s Guide 79
Features and Functions 2
Remote Interface Operation: The following command selects the trigger slope for
the external trigger (Ext Trig) connector:
TRIGger:SLOPe {POSitive>NEGative}
The following query returns the selected trigger slope for the Ext Trig connector:
TRIGger:SLOPe?
This query returns POS or NEG.
The following command selects the signal slope for the voltmeter complete (VM
Comp) signal:
OUTPut:TRIGger:SLOPe {POSitive>NEGative}
The following query returns the selected trigger slope for the VM Comp connector:
OUTPut:TRIGger:SLOPe?
This query returns POS or NEG.
Refer to the Keysight 34410A/11A/L4411A Programmers Reference Help for
a complete description and syntax for this command.