Testing Performance 2
6000 X-Series Oscilloscopes Service Guide 23
To test digital channels (MSO models only)
The acquisition system testing provides confidence that the acquisition system is
functioning correctly. It does not, however, check a particular specification.
1 Disconnect all probes from the circuit under test and from any other input
source.
2 Using probe leads and grabbers, connect digital channels D0, D1, D2, and D3
to the Probe Comp signal on the center of the front panel.
3 Press the [AutoScale] key.
If four square waves appear, the acquisition system is functioning correctly.
If the square waves do not appear, go to the “Troubleshooting” chapter. Then
return here to finish testing the digital channels.
4 Disconnect the digital channels from the calibration point.
5 Use steps 2 and 3 to test the following sets of digital channels. After you test
one set of digital channels, remove them before connecting the next set.
D4, D5, D6, D7
D8, D9, D10, D11
D12, D13, D14, D15

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies InfiniiVision 6000 X Series and is the answer not in the manual?

Keysight Technologies InfiniiVision 6000 X Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelInfiniiVision 6000 X Series
CategoryTest Equipment
LanguageEnglish

Related product manuals