Page #81 background image
Troubleshooting 4
6000 X-Series Oscilloscopes Service Guide 81
Table 22 Power supplies to front panel board (AutoProbe power, from power board J106)
Supply Test Point Expected Value If Good If Bad
VP13VF_PRB J106, pin 1 +13 V ±2% Supply is good. Replace the power board.
VM13VF_PRB J106, pin 2 -13 V ±2% Supply is good. Replace the power board.
VM6V5_PRB J106, pin 3 -6.512 ±3% Supply is good. Replace the power board.
VP6V5_PRB J106, pin 4 +6.512 ±3% Supply is good. Replace the power board.