42 6000 X-Series Oscilloscopes Service Guide
2 Testing Performance
16 Move the power splitter from the channel 1 to the channel 2 input.
17 Turn off the current channel and turn on the next channel using the channel
keys.
18 Repeat steps 3 through 17 for the remaining channels, setting the parameters
of the channel being tested where appropriate.