5 Maintenance
162 Keysight U1253B User’s and Service Guide
Table 5-5 Basic troubleshooting procedures
Malfunction Troubleshooting procedure
No OLED display after switching ON Check battery. Charge or replace battery.
No beeper tone
Check the Setup mode to verify whether the beeper function has been set to OFF.
If so, select the desired driving frequency.
Failed to measure current Check the fuse.
No charging indication
[a]
Check 440 mA fuse.
Check external DC adapter to ensure that its output is DC 24 V and that the plugs
are inserted properly into the charging terminals.
Battery life very short after full charge/
Battery not able to charge after pro-long
storage period
Check if the correct rechargeable battery is used.
Check if the correct nominal voltage level (7.2 V or 8.4 V) is selected in the battery
setting at Setup Mode.
Try to charge and discharge for 2 or 3 cycles in order to maintain the battery’s
highest capacity.
NOTE: The performance of the rechargeable battery may degrade with time.
Remote control failure
The Keysight logo on the IR-USB cable connected to the multimeter should be
facing up.
Check the baud rate, parity, data bit, and stop bit (default is 9600, None, 8, and 1)
in the Setup mode.
Ensure that the required driver for IR-USB has been installed.
Notes for basic troubleshooting procedures table:
[a] Never turn the rotary switch of the multimeter from the OFF position when it is charging.