6 Performance Tests and Calibration
192 Keysight U1253B User’s and Service Guide
Calibration From Front Panel
Calibration process
The following general procedure is the recommended method to complete a full
instrument calibration.
1 Read and implement “Test Considerations” on page 172.
2 Perform the verification tests (refer to Table 6-2 on page 174) to characterize
the instrument.
3 Perform the calibration procedures (refer to the “Calibration procedures” on
page 193; see also Adjustment Considerations” on page 187).
4 Secure the instrument after calibration.
5 Take note of the new security code (if it has been changed) and the calibration
count in the instrument maintenance records.
NOTE
Make sure to quit the adjustment mode before switching off the instrument.