Interval logging
Firstly ensure that interval (time) logging is specified in Setup mode.
1 Press for more than 1 second to store the present value and function on
the primary display into the non-volatile memory. The and the logging
index will be indicated. Subsequent readings are automatically logged into the
memory at the interval (LOG TIME) specified in Setup mode. Refer to
Figure 3-10 on page 100 for how to operate this mode.
2 Press for more than 1 second to exit this mode.
NOTE
The maximum number of readings that can be stored is 1000 entries. When the
1000 entries are all occupied, the logging index will indicate “Full”.
NOTE
When interval (time) logging is running, all keypad operations are disabled,
except for Log, which, when pressed for longer than 1 second, will exit this
mode. Furthermore, Auto Power Off is disabled during interval logging.