EasyManua.ls Logo

Keysight Technologies U1253B - Page 190

Keysight Technologies U1253B
205 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
6 Performance Tests and Calibration
190 Keysight U1253B User’s and Service Guide
AC mA/AC A 50 mA 3.000 mA (1 kHz) 0.9 to 1.1 × reference input value
30.000 mA (1 kHz) 0.9 to 1.1 × reference input value
500 mA 30.00 mA (1 kHz) 0.9 to 1.1 × reference input value
30.000 mA (1 kHz) 0.9 to 1.1 × reference input value
5 A 0.3000 A (1 kHz) 0.9 to 1.1 × reference input value
3.0000 A (1 kHz) 0.9 to 1.1 × reference input value
10 A 0.3000 A (1 kHz) 0.9 to 1.1 × reference input value
10.000 A (1 kHz) 0.9 to 1.1 × reference input value
Capacitance Open OPEN Open terminals
10 nF 3.000 nF 0.9 to 1.1 × reference input value
10.000 nF 0.9 to 1.1 × reference input value
100 nF 10.00 nF 0.9 to 1.1 × reference input value
100.00 nF 0.9 to 1.1 × reference input value
1000 nF 100.0 nF 0.9 to 1.1 × reference input value
1000.0 nF 0.9 to 1.1 × reference input value
10 μF 10.000 μF 0.9 to 1.1 × reference input value
100 μF 100.00 μF 0.9 to 1.1 × reference input value
1000 μF 1000.0 μF 0.9 to 1.1 × reference input value
10 mF 10.000 mF 0.9 to 1.1 × reference input value
Table 6-3 Valid adjustment reference input values (continued)
Function Range Reference input value Valid range for reference input

Table of Contents

Related product manuals