Performance Tests and Calibration 6
Keysight U1253B User’s and Service Guide 197
AC mV 50 mV 3.000 mV (1 kHz)
30.000 mV (1 kHz)
30.000 mV (10 kHz)
3 mV, 1 kHz
30 mV, 1 kHz
30 mV, 10 kHz
500 mV 30.00 mV (1 kHz)
300.00 mV (1 kHz)
300.00 mV (10 kHz)
30 mV, 1 kHz
300 mV, 1 kHz
300 mV, 10 kHz
1000 mV 300.0 mV (1 kHz)
1000.0 mV (1 kHz)
1000.0 mV (10 kHz)
(done)
300 mV, 1 kHz
1000 mV, 1 kHz
1000 mV, 10 kHz
Resistance
[d]
Short SHORT Dual banana shorting plug with
copper wire
50 MΩ OPEN
10.000 MΩ
Unplug all test leads or shorting plug,
and leave the terminals open
10 MΩ
5 MΩ 3.0000 MΩ 3 MΩ
500 kΩ 300.00 kΩ 300 kΩ
50 kΩ 30.000 kΩ 30 kΩ
5 kΩ 3.0000 kΩ 3 kΩ
500 Ω 300.00 Ω
(done)
300 Ω
Diode Short SHORT Dual banana shorting plug with
copper wire
2 V 2.0000 V (done) 2 V
Table 6-4 List of calibration items (continued)
Function Range
Calibration item
[a]
Reference input

Table of Contents

Related product manuals