Features and Functions 4
Keysight 34970A/34972A User’s Guide 115
– You can specify a scan count which sets the number of alarms that will be
allowed before terminating the scan. See “Scan Count” on page 117 for more
information.
– All readings from the scan are stored in non-volatile memory. Readings
accumulate in memory until the scan is terminated (until the scan count is
reached or until you abort the scan).
– Mx+B scaling and alarm limits are applied to measurements during the scan
and all data is stored in non-volatile memory.
– Front-Panel Operation:
SCAN ON ALARM
To enable the Monitor function select the desired channel and then press
. To initiate the scan, press . When an alarm event occurs, the scan
starts and readings are stored in memory.
– Remote Interface Operation: The following program segment configures the
instrument to scan when an alarm occurs
.
To stop a scan, press and hold .