6 53220A/53230A Math, Graphing, and Data Logging
212 Keysight 53220A/53230A User’s Guide
Inverting the reading (1/x) Each reading as it is taken can be inverted prior to
its use in the Mx-B equation which thus becomes M/x-B. Inversion is enabled with
the command:
CALCulate[1]:SCALe:INVert {OFF|ON}
CALCulate[1]:SCALe:INVert? (query form)
ON - inverts the reading. OFF disables reading inversion.
Following a reset (*RST) or instrument preset (SYSTem:PRESet) reading inversion
is disabled.
Assigning reading units
NOTE
When Invert x is enabled (On) the front panel reading units (Hz, sec) are turned
off. See “Assigning reading units” if displayed units are required.